Using patented tunable diode laser absorption spectroscopy (TDLAS) technology, the SS2100 gas analyzer measures specific gases (H2O, CO2, H2S, NH3, and C2H2) with precision while avoiding interferences that are common with traditional infrared analyzers. The SS2100 analyzer requires little maintenance and does not need recalibration or periodic replacement parts.
Gas analysis: ppb level measurement of H2O & NH3, ppm level measurement of H2O, H2S, CO2, NH3, and C2H2, and percent (%) level measurement of H2S and CO2
Natural gas: production, storage, transportation, distribution
Check out the findings from the GERG report on the use of a Raman Rxn4 analyzer and Raman Rxn-41 cryogenic probe for the reliable chemical composition measurement in LNG custody transfer applications.
H2S (Hydrogen Sulfide): 0 to 10 ppmv 0 to 1000 ppmv 0 to 5000 ppmv 0 to 5% volume
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Channels
1
Analyte and Measurement ranges
H2O (Moisture): 0 to 10 ppmv 0 to 100 ppmv 0 to 50 ppmv 0 to 5000 ppmv
Measured Variables
Concentration Moisture Dew Point Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Channels
1
Analyte and Measurement ranges
CO2 (Carbon Dioxide): 0 to 10 ppmv 0 to 1000 ppmv 0 ppmv to 5000 ppmv 0 to 5% volume 0 to 10% volume 0 to 20% volume
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC
Communication
Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs: 2, Flow Alarm, Validation Request
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Channels
2
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0 to 10 ppmv (minimum) 0 to 1000 ppmv (maximum) H2O (Moisture): 0 to 100 ppmv (minimum) 0 to 5000 ppmv (maximum)
Measured Variables
Concentration Moisture Dew Point Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC
Communication
Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O) Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs (H2S): 2, Flow Alarm, Validation Request Digital Outputs (H2O): 2, Concentration Alarm, General Fault
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Channels
2
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0 to 10 ppmv (minimum) 0 to 1000 ppmv (maximum) CO2 (Carbon Dioxide): 0 to 5000 ppmv 0 to 5% volume 0 to 10% volume 0 to 20% volume
Measured Variables
Concentration Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC
Communication
Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (CO2) Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs (H2S): 2, Flow Alarm, Validation Request Digital Outputs (CO2): 2, Concentration Alarm, General Fault
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Hazardous area approvals
CSA Class I, Division 2 CSA Class I, Zone 2
Degree of protection
IP66, Type 4X
Measuring principle
TDLAS
Channels
3
Analyte and Measurement ranges
H2S (Hydrogen Sulfide): 0 to 10 ppmv (minimum) 0 to 1000 ppmv (maximum) H2O (Moisture): 0 to 100 ppmv (minimum) 0 to 5000 ppmv (maximum) CO2 (Carbon Dioxide): 0 to 5000 ppmv 0 to 5% volume 0 to 10% volume 0 to 20% volume
Measured Variables
Concentration Moisture Dew Point Cell Pressure Cell Temperature
Ambient Temperature range
-20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F)
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)
Analyzer Wetted materials
316L stainless steel FKM O-Rings Glass
Power supply
120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W (18 to 24 VDC - optional for electronics only) 3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC
Communication
Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O/CO2) Protocol: Modbus Gould RTU or Daniel RTU or ASCII Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active Digital Inputs (H2S): 2, Flow Alarm, Validation Request Digital Outputs (H2O/CO2): 2, Concentration Alarm, General Fault
Housing materials
Electronics: 304 or 316L stainless steel Sample System Enclosure: 304 or 316L stainless steel
Understanding factory validation gas - Verify TDLAS performance
PU (Publication)
English version - 02/2025
New version available in English
An overview about the validation of TDLAS performance and the determination of proper validation gas backgrounds for the support and service of TDLAS analyzers (SS2100, SS2100i, J22, JT33, SS500).
Real-time quality measurements in biogas and biomethane
PU (Publication)
English version - 02/2025
New version available in English
A discussion on using TDLAS / QF analyzers (SS500, J22, JT33, SS2100, and OXY5500) to measure the concentration of compounds in biogas and biomethane in the Oil & Gas industry.
An overview about the engineered approval drawings offered for TDLAS and QF analyzers (SS2100, SS2100i, J22, JT33, OXY5500, SS500) to enable customers to confirm that the system fabrication meets their approval.
TDLAS and QF analyzers application guide - Accurate and
reliable measurement of H2O, H2S, CO2, and O2
PU (Publication)
English version - 02/2025
New version available in English
A guide to TDLAS and QF technology application notes (natural gas, hydrogen, biomethane, carbon capture, LNG, gas processing, NGL fractionation, petrochemical, syngas, ammonia, and refining) for the oil & gas and chemical industries (relating to J22, JT33, SS500, SS2100, SS2100i, and OXY5500 analyzers).
Process analyzer solutions - Expertise for optimal performance
Competence Brochure (CP)
English version - 12/2024
New version available in English
An overview brochure about process analysis solutions offered by the Endress+Hauser Process Analyzer Support team in Europe, including Raman spectroscopy, TDLAS, and QF technology-based solutions.
Analysis system for quality control of biomethane - Reducing carbon
footprints with renewable energy
Competence Brochure (CP)
English version - 11/2023
New version available in English
An overview of Endress+Hauser's TDLAS/QF measurement systems (J22, SS2100 and SS2100i TDLAS gas analyzers, OXY5500 optical oxygen QF analyzer) for quality control of biomethane for the oil & gas industry which helps companies reduce emissions using renewable energy.
TDLAS analyzers for olefins production - Accurate and
reliable measurement of H2O, H2S, and CO2
Innovation (IN)
English version - 02/2025
New version available in English
An overview brochure of our tunable diode laser absorption spectroscopy (TDLAS) gas analyzers (including the JT33, SS2100, SS2100i, and TDLAS gas analyzers) to measure water / moisture (H2O), hydrogen sulfide (H2S), and carbon dioxide (CO2) in the olefins / chemical industry using SpectraSensors technology.
Optical analysis technologies for LNG - Reliable, real-time TDLAS
and Raman measurement
Innovation (IN)
English version - 01/2025
New version available in English
A portfolio overview brochure about our TDLAS and Raman spectroscopy analyzer systems for the LNG / Oil & Gas industry (includes the SS2100, SS2100i, J22, JT33, Raman Rxn4 analyzer, and the Raman Rxn-41 probe).
TDLAS analyzers for refinery gases - Accurate and reliable
measurement of H2S and H2O
Innovation (IN)
English version - 10/2023
New version available in English
An overview brochure of our tunable diode laser spectroscopy (TDLAS) gas analyzers (including the JT33, SS2100, and SS2100i TDLAS gas analyzers) to measure water / moisture (H2O) and hydrogen sulfide (H2S) in the refining industry segment of the chemical industry / Oil & Gas industry using SpectraSensors technology.
TDLAS and QF analyzers technology guide - Principle of operation,
configurations, and certification information
Innovation (IN)
English version - 02/2025
New version available in English
A technology guide examining our tunable diode laser absorption spectroscopy (TDLAS) and quenched fluorescence (QF) analyzer portfolio (including the J22, JT33, SS500, SS2100, and SS2100i TDLAS gas analyzers, and the OXY5500 precision oxygen analyzer) to measure water / moisture (H2O), hydrogen sulfide (H2S), carbon dioxide (CO2), and oxygen (O2) in natural gas process streams within the chemical industry / Oil & Gas industry using SpectraSensors technology.